DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
4 - 20 Section 4-2 - General Procedures
4-2-10-2 Loading Presets from an MO Disk
NOTE: It is not possible to move presets between units with different software versions.
NOTE: You can successfully copy presets from one V7 to another, provided they run the same software
version.
Table 4-4 Loading Presets
STEP TASK FUNCTION
1.
Insert the MO disk with the archived
presets into the MOD.
2.
Go to Config (F2) > Admin > Restore
The Restore window is displayed.
3.
Select the Check box “User defined
Configuration”.
4.
Select the MOD as source.
5.
Press RESTORE NOW The Presets are restored to the System.
6.
Press ALT + E on the alphanumeric
keyboard.
The MO disk is ejected.