4 - 38 Section 4-3 - Functional Checks
4-3-6-8 Check DDP, Diff, Power and Dynamic Range
 
Table 4-11    2D Mode: Check DDP, Diff, Power and Dynamic Range
Step Task Expected Result(s)
1.
Use the vertical arrows to select DDP (Data 
Dependent Processing).
Use the horizontal arrows to adjust DDP.
Performs temporal processing which reduces random 
noise without affecting the motion of significant tissue 
structures. 
An index number is displayed in the status window 
(under Proc) to indicate the relative DDP level.
2.
Use the vertical arrows to select Diff.
Use the horizontal arrows to turn Diff On/Off.
Affects the level of reverberations in the image.
When turned ON, the frame rate (or the number of 
focal zones) will decrease, while the reverberations 
will be attenuated.
3.
Use the vertical arrows to select Power.
Use the horizontal arrows to adjust Power.
Controls the amount of acoustic power applied in all 
modes.
When power is set to maximum, it is equal to or less 
than the maximum acoustic power permitted by the 
FDA.
The Thermal Index (TI) and the Mechanical Index (MI) 
are displayed on the screen.
4.
Use the vertical arrows to select Dynamic Range. 
Use the horizontal arrows to adjust Dynamic Range.
Enables control of the dynamic range or contrast of 
the image.
When Dynamic Range is set high, the image is softer 
and more low-level data is visible.