DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
4 - 56 Section 4-3 - Functional Checks
4-3-10 4D Volume Rendering Test (cont’d)
5.) Using the trackball rotate and translate the volume rendering to display the structure of interest as
described in the section below. Press Angle to get a pre-defined orientation before translating and/
or rotating the volume rendering. If necessary, press Clear to reset the orientation to the default
position.
6.) Optimize the volume rendering using 4D Gain, Shading and Smoothness controls.
7.) If desired, zoom in the area of interest.
8.) If desired, press Layout to display the volume rendering in a single screen.
9.) If necessary, use Crop tool to remove unwanted structures.
10.)Select Img. Store to save the acquisition.