DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 5 - Components and Functions (Theory) 5 - 43
5-3-16-10 Test Points for DC Voltages
See Figure 5-29 "FEC-2, Test points for DC Voltages" on page 5-42.
5-3-16-11 Fuses and Dip-switches
None
Table 5-34 Test points for DC Voltages
TEST POINT DC VOLTAGE ACCEPTABLE RANGE
TP 8
+5 V +4.75 V to +5.25 V
TP 10
+15 V +/- 10%
TP 9
-15 V +/- 10%
TP 7
-5 Va +/- 10%
TP 20
+3.3 Va +/- 10%
TP 17
+6 Va +/- 10%
TP 21
+10 Va +/- 10%