DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 5 - Components and Functions (Theory) 5 - 131
5-11-2-3 Distribution of Voltages via AC Controller
The 20 VAC is routed via the fan regulator to the system fan.
Both 115 VAC and 230 VAC are routed via fuses.
115 VAC is routed to the Monitor outlet (MONITOR)
An AC Voltage selector is used to select the correct voltage for the Peripherals. The selected
voltage (230 VAC or 115 VAC) is routed to the Peripherals AC outlet (PERIPH’S)
230 VAC is distributed to;
- Modem AC outlet (MODEM)
- Back-End Processor AC outlet (PC)
- Front End Card Cage AC outlet (RACK)
Figure 5-84 Connector Pin-out - AC Connectors on AC Controller