EasyManua.ls Logo

GE Vivid 7 - Page 486

GE Vivid 7
858 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
5 - 140 Section 5-11 - Power Distribution
5-11-4-5 Outputs - FB200574
5-11-4-6 Fuses
Output Power Protection Circuit with these functions:
- Protect probe and TX board if a failure make the voltage exceed limits
- Protect the power supply if there is a failure in probe or transmitter circuit
A watch-dog disables the voltage (set it to zero Volt) if the;
- serial communication with the FEC-2 module stops
- nominal Power limit is exceeded
- maximum Power limit is exceeded
5-11-4-7 Jumpers and Dip-switches
None.
5-11-4-8 LEDs
None.
5-11-4-9 Power Loss
A power loss may be due to:
The Mains Switch has been switched to OFF
The Mains cable has been disconnected
Brown-out
If a power loss occur, all AC power distribution within the unit is lost. The Front-End Card Rack stops
functioning, the peripherals and the monitor also looses its power.
On units with an internal UPS, the Back-End Processor automatically switch over to the internal battery,
and the Back-End Processor starts an automatic power down sequence.
Table 5-92 Output Voltages, Plug P2
OUTPUT DESCRIPTION CONNECTED TO:
TSV1
TX Voltage 1
Voltage can be programmed to vary from +/- 2.5 V to +/- 80V
Used to drive the transmitters in 2D Mode and M mode
TX board
TSV2
TX voltage 2
Voltage can be programmed to vary from +/- 2,5 V to +/- 40V
Used to drive the transmitters in Doppler (CW/PW) mode
and in Color Flow Mapping (CFM) mode
TX board
PMXVOUT
Probe Multiplexing Voltage
A fixed +/-100 V Voltage
Used for the multiplexers in linear array probes with more
than 128 transducer elements
RLY board

Table of Contents

Related product manuals