GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 6 - Service Adjustments 6 - 11
6-8-2-4 Direction Lock Adjustment (cont’d)
5.) Turn adjustment nuts until the balljoints matches.
6.) Press the balljoint together.
7.) Install the lock pins, see Figure 6-5.
8.) Fasten the retaining nuts.
9.) Test Direction Lock.
Figure 6-6 Balljoint with Lock Pin
Adjustment nut
Retaining nut
Lock Pin