EasyManua.ls Logo

GE Vivid 7 - Page 566

GE Vivid 7
858 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
7 - 54 Section 7-7 - Common Diagnostics
7-7-2-3 Memory Tests
Memory Test (All) - Eighteen test patterns are used to test memory locations. If errors are
detected, the address of the block the error was detected in is shown. During testing, these
diagnostics will also look out for parity errors and other exceptions. These will be added to the Test
Log as errors if encountered.
Memory: Pattern Test - The program uses several test patterns to test as much memory as
possible.
Memory: Parity Test - This test checks for parity errors on the memory bus during intensive string
transfer operations. If an error is detected, the address of the failed block is displayed.
Figure 7-56 Memory Tests