GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
8 - 28 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x
8-7-12-4 Loading Application Software v7.x on Vivid 7 with BEP4.3 (cont’d)
4.) Select “C. Reload Application SW"
5.) Confirm the selection by pressing Y.
The system reboots and the Application software installation starts automatically.
When completed, the Vivid 7 reboots and starts up normally. The User Presets, the System
Configurations and the local Patient Archive remains as before the software reload.
Figure 8-24 System Recovery Menu
NOTICE
If you select “A” in the next step, ALL existing software and data will be erased. If backup has
not been performed as described earlier in this manual; ALL DATA like Patient Database,
System Configuration and User Configurations (Customer Presets) will be lost.
System Recovery Menu
A. Full reload (erase all partitions)
B. Reload Partition C:\ only (Keep Archive)
C. Reload Application Software
R. Restart System
Select (A, B, C,,R) _