GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
8 - 32 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x
8-7-12-6 Application SW Loading - Software Version v7.x/v6.x/v5.x/v4.x
1.) Insert the applicable CD labelled Vivid 7 & EchoPAC PC Application SW.
2.) Keep the ON/Standby switch on the operator panel pressed for a few seconds to power down the
scanner. (This is called Forced Shut Down).
3.) Wait until the LCD’s on the Operator Panel is switched OFF and the ON/Standby Switch on the
Operator Panel has turned yellow.
4.) Power ON the System, see 4-2-2 "Power ON/ Boot UP" on page 4-3.
5.) A dialogue box, called Start Application, will appear, select Install SW.
6.) A dialogue box, called Startloader, will appear, select OK to continue.
7.) A dialogue box will appear, also named Startloader stating “New System SW installation will start
now”, select OK to continue.
8.) A Command prompt window will appear, hit any key on the alphanumeric keyboard to continue.
9.) The messages: “Unpacking SW. This will take several minutes.” and “Installing new application
software. This will take several minutes” will be displayed on the screen.
10.)Several dialogue boxes will appear on the screen during the SW load.
11.)Remove the CD from the CD/DVD drive when prompted.
12.)Press any key to continue.
13.)The Scanner will now power down and reboot.
WARNINGWARNING
Do not interrupt the software loading at any time.