EasyManuals Logo

Keysight Technologies 6811C Programming Guide

Keysight Technologies 6811C
315 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #102 background imageLoading...
Page #102 background image
3 Language Dictionary
102 Keysight Models 6811C, 6812C, and 6813C Programming Guide
MEASure:VOLTage:RANKed?
FETCh:VOLTage:RANKed?
Keysight 6811C/6812C/6813C Option 020 only.
Scans through the buffer of 4096 instantaneous voltage measurements, and
returns the voltage value that corresponds to the percentile rank given. Specifying
a percentile of 0 returns the lowest value in the buffer. A percentile of 100 returns
the highest value in the buffer. A percentile of 50 returns the median value of the
buffer. Stated another way, if you specify a percentile of 60, 60 percent of the
voltage readings in the buffer are less that the returned value, and 40 percent of
the readings in the buffer are greater than the returned value.
If the voltage waveform consists of rectangular pulses that have narrow
overshoots, using percentiles of approximately 5 and 95 (depending on the width
of the pulses and the width of the overshoots) is a good way to determine the
voltages at the flat portions of the pulses.
Query syntax MEASure:[SCALar]:VOLTage:RANKed? <percentile>
FETCh:[SCALar]:VOLTage:RANKed? <percentile>
Parameters 0 to 100
Examples
MEAS:VOLT:RANK? 50 FETC:VOLT:RANK? 5
Returned parameters <NR3>

Table of Contents

Other manuals for Keysight Technologies 6811C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 6811C and is the answer not in the manual?

Keysight Technologies 6811C Specifications

General IconGeneral
BrandKeysight Technologies
Model6811C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals