Language Dictionary 3
Keysight Models 6811C, 6812C, and 6813C Programming Guide 71
CALibrate:VOLTage:OFFSet
CALibration:VOLTage:OFFSet
This command can only be used in calibration mode. It initiates the calibration of
the offset voltage programming circuits.
CALibrate:VOLTage:PROTection
CALibration:VOLTage:PROTection
This command can only be used in calibration mode. It calibrates the overvoltage
protection (OV) circuit. The command automatically performs the calibration.
CALibrate:SAVE is required to store the new OV constant in non-volatile memory.
CALibrate:VOLTage:PROTection is a sequential command that takes several
seconds to complete.
Command syntax CALibrate:VOLTage:OFFSet
CALibration:VOLTage:OFFSet
Parameters None
Examples
CAL:VOLT:OFFS
Related commands CAL:SAVE CAL:STAT CAL:LEV
Command syntax CALibrate:VOLTage:PROTection
CALibration:VOLTage:PROTection
Parameters None
Examples
CAL:VOLT:PROT
Related commands CAL:SAVE CAL:STAT