INITiate:SEQuence
INITiate:NAME
The INITiate commands control the initiation of both the transient generator and
the measurement trigger systems. They cause the trigger system to make a
transition from the Idle state to the Waiting-for-Trigger state. If the trigger system
is not in the Idle state, the initiate commands are ignored.
INITiate:SEQuence and INITiate:NAME initiate the trigger systems to reference
trigger sequences. INITiate:SEQuence references a trigger sequence by its
number, while INITiate:NAME references a sequence by its name. The
correspondence between sequence names and numbers is:
Sequence number Sequence number Description
1 (the default) TRANsient Step, pulse, or list transient trigger sequence
3 ACQuire Measurement acquire trigger sequence
Command syntax INITiate[:IMMediate]:SEQuence[1 | 3]
INITiate[:IMMediate]:NAME<name>
Parameters For INIT:NAME: TRANsient | ACQuire
Examples
INIT:SEQ1 INIT:NAME ACQ
Related commands ABOR INIT:CONT TRIG *TRG