3 Language Dictionary
114 Keysight Models 6811C, 6812C, and 6813C Programming Guide
SENSe:NSELect
Keysight 6811C/6812C/6813C Option 020 only.
Selects the measurement source that will return data when a query is sent. A
parameter value of 1 selects the Main Output, which measures the actual output
voltage and current of the ac source. A value of 2 selects the Power Analyzer
input, which measures the voltage and current at the V SENSE and I SENSE input
channels.
Only the following commands are affected by the measurement source selection:
– Queries beginning with MEASure or FETCh
– SENSe:CURRent:ACDC:RANGe
SENSe:SWEep:OFFSet:POINts
This command defines the trigger point relative to the start of the returned data
record when an acquire trigger is used. The values can range from –4095 to 2E9.
When the values are negative, the values in the beginning of the data record
represent samples taken prior to the trigger.
Command syntax SENSe:NSELect <NR1>
Parameters 1 | 2
*RST value 1
Examples
SENS:NSEL 2
Query syntax SENSe:NSELect?
Returned parameters <NR1>
Command syntax SENSe:SWEep:OFFSet:POINts <NRf+>
Parameters –4096 through 2E9 | MINimum | MAXimum
*RST value 0 (zero)
Examples
SENS:SWE:OFFS:POIN -2047
Query syntax SENSe:SWEep:OFFSet:POINts?
Returned parameters <NR3>
Related commands SENS:SWE:TINT MEAS:ARR

Table of Contents

Other manuals for Keysight Technologies 6811C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 6811C and is the answer not in the manual?

Keysight Technologies 6811C Specifications

General IconGeneral
BrandKeysight Technologies
Model6811C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals