3 Language Dictionary
130 Keysight Models 6811C, 6812C, and 6813C Programming Guide
FUNCtion
This command selects the shape of the output voltage waveform as follows:
– SINusoid: A sine wave is the output.
– SQUare: A square wave is the output.
– CSINusoid: The output is a clipped sine wave. Both positive and negative peak
amplitudes are clipped at a value determined by the FUNCtion:CSINusoid
command.
– <table>: The output shape is described by one of the user-defined waveform
tables
The maximum peak voltage that the ac source can output is 425 V
peak
. This
includes any combination of voltage, voltage offset, and function shape values.
Therefore, the maximum value that can be programmed depends on the
peak-to-rms ratio of the selected waveform. For a sine wave, the maximum
voltage that can be programmed is 300 V
rms
.
Command syntax [SOURce:]FUNCtion[:SHAPe][:IMMediate] <shape>
Parameters SINusoid | SQUare | CSINusoid | <table>
*RST value SINusoid
Examples
FUNC SIN FUNC:SHA SQU
Query syntax [SOURce:]FUNCtion[:SHAPe][:IMMediate]?
Returned parameters <CRD>
Related commands FUNC:MODE FUNC:TRIG VOLT

Table of Contents

Other manuals for Keysight Technologies 6811C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 6811C and is the answer not in the manual?

Keysight Technologies 6811C Specifications

General IconGeneral
BrandKeysight Technologies
Model6811C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals