OUTPut
This command enables or disables the ac source output. The state of a disabled
output is an output voltage amplitude set to 0 volts, with output relays opened.
The query form returns the output state.
For Keysight 6812C/6813C Option 026 only, this command controls the on/off
state of both the main and auxiliary outputs.
OUTPut:COUPling
This command enables AC or DC output coupling. When the output coupling is
set to AC, a DC leveling loop attempts to maintain zero average output voltage.
The loop has a corner frequency of about 2 Hz. It will not prevent short transient
waveforms that may have non-zero average voltage, but will cause a settling
transient to an average value of 0 volts.
The output coupling must be set to DC to obtain DC output with VOLTage:OFFSet,
or to generate output transients that have net DC components.
Command syntax OUTPut[:STATe] <bool>
Parameters 0 | 1 | OFF | ON
*RST value OFF
Examples
OUTP 1 OUTP:STAT ON
Query syntax OUTPut[:STATe]?
Returned parameters 0 | 1
Related commands *RCL *SAV
Command syntax OUTPut:COUPling <coupling>
Parameters AC | DC
*RST value AC
Examples
OUTP:COUP DC
Query syntax OUTPut:COUPling?
Returned parameters <CRD>
Related commands *RCL *SAV

Table of Contents

Other manuals for Keysight Technologies 6811C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 6811C and is the answer not in the manual?

Keysight Technologies 6811C Specifications

General IconGeneral
BrandKeysight Technologies
Model6811C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals