3 Language Dictionary
172 Keysight Models 6811C, 6812C, and 6813C Programming Guide
Bit configuration of Questionable Status registers
STATus:QUEStionable?
This query returns the value of the Questionable Event register. The Event register
is a read-only register which holds (latches) all events that are passed by the
Questionable NTR and/or PTR filter. Reading the Questionable Event register
clears it.
Bit position 15141312111098 – 543210
Bit name n.u.
Meas
Ovld
n.u.
CL
rms
Rail
CL
peak
RI n.u. OT UNR SOA OCP OV
Bit weight 16384 4096 2048 1024 512 16 8421
n.u. = Not used
OV = Overvoltage protection has tripped
OCP = Overcurrent protection has tripped
SOA = Safe operating area protection has tripped
UNR = Output is unregulated
OT = Overtemperature protection has tripped
RI = Remote inhibit is active
CL peak = Peak current limit is active
Rail = Rail protection tripped
CL rms = rms current limit is active
MeasOvld = Current measurement exceeded low current range capability
Query syntax STATus:QUEStionable[:EVENt]?
Parameters None
Examples
STAT:QUES:EVEN?
Returned parameters <NR1> (register value)
Related commands *CLS STAT:QUES:NTR STAT:QUES:PTR