Language Dictionary 3
Keysight Models 6811C, 6812C, and 6813C Programming Guide 127
Source Subsystem (Frequency Modulation) Commands
Keysight 6811C/6812C/6813C Option 020 only.
The following SCPI commands are used to program the output frequency of the ac
source. These commands apply only to the Main Output of the ac source and do
not affect the measurement capability of the unit.
Subsystem syntax
FM
Keysight 6811C/6812C/6813C Option 020 only.
Turns off or on the frequency modulation of the Main Output. Frequency
modulation operates only if FREQuency:MODE is set to FIXed.
[SOURce:]
FM
[:STATe] <bool> Turns off or on the frequency modulation of the Main Output
:DEViation <n> Sets the peak frequency deviation in hertz
:FREQuency <n> Sets the modulating frequency in hertz
Command syntax [SOURce:]FM[:STATe] <bool>
Parameters 0 | 1 | OFF | ON
*RST value OFF
Examples
FM 1 FM:STATE ON
Query syntax [SOURce:]FM[:STATe]?
Returned parameters 0 | 1
Related commands FM:DEV FM:FREQ