Language Dictionary 3
Keysight Models 6811C, 6812C, and 6813C Programming Guide 133
FUNCtion:CSINusoid
This command sets the clipping level when a clipped sine output waveform is
selected. The clipping characteristics can be specified in two ways:
– The clipping level is expressed as a percentage of the peak amplitude at which
clipping occurs. The range is 0 to 100 percent. These are the default units
when the optional THD suffix is not sent.
– The clipping level is expressed at the percentage of total harmonic distortion in
the output voltage. The range is 0 to 43 percent. The optional THD suffix is
sent to program in these units.
Command syntax [SOURce:]FUNCtion[:SHAPe]:CSINusoid <Nrf> [THD]
Parameters 0 to 100% | 0 to 43% THD
*RST value 100% | 0% THD (no clipping)
Examples
FUNC:CSIN 80 FUNC:CSIN 10 THD
Query syntax [SOURce:]FUNCtion[:SHAPe]:CSINusoid? [THD]
Returned parameters <NR3>
Related commands FUNC FUNC:MODE