3 Language Dictionary
188 Keysight Models 6811C, 6812C, and 6813C Programming Guide
SYSTem:COMMunicate:RLSTate
Sets the interface to the following states:
Local state: Enables the front panel controls.
Remote state: Disables all front panel controls except the Local Key. Pressing
the Local Key while in the Remote state returns the front panel to the Local
state
Remote-lockout state: Disables all front panel controls including the Local
key. Use SYSTem:LOCal or SYSTem:COMMunicate:RLSTate LOCal to return
the front panel to the Local state.
SYSTem:COMMunicate:TCPip:CONTrol?
Reads the initial control connection port number for sockets communications.
This connection is used to send and receive commands and queries. Use the
control-socket connection to send a Device Clear (DCL) to the ac source or to
detect pending SRQ events.
Command syntax SYSTem:COMMunicate:RLSTate <mode>
Parameters LOCal | REMote | RWLock
Examples
SYST:COMM:RLST LOC
Query syntax SYSTem:COMMunicate:RLSTate?
Returned parameters <CRD>
Related commands SYST:LOC SYST:REM SYST:RWL
NOTE
Refer to the Keysight Models 6811C, 6812C, and 6813C User’s and Service
Guide under “Using Sockets” for more information
Query syntax SYSTem:COMMunicate:TCPip:CONTrol?
Parameters None
Examples
SYST:COMM:TCP:CONT
Returned parameters <NR1>
Related commands SYST:COMM:LAN:CONT

Table of Contents

Other manuals for Keysight Technologies 6811C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 6811C and is the answer not in the manual?

Keysight Technologies 6811C Specifications

General IconGeneral
BrandKeysight Technologies
Model6811C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals