Keysight Models 6811C, 6812C, and 6813C Programming Guide 67
CALibrate:SHUNt
CALibration:SHUNt
Keysight 6811C/6812C/6813C Option 020 only.
Allows you to enter the external current shunt value (in ohms). The external
current shunt is used for external current measurement. The programming range
is 1E-6 to 1E6. This parameter is non-volatile, so there is no *RST value. As with
other CAL commands, calibration must first be enabled with CAL:STATe, and the
value must be made permanent with CAL:SAVE.
Command syntax CALibrate:SHUNt <NRf+>
CALibration:SHUNt <NRf+>
Parameters 1E-6 to 1E6 | MINimum | MAXimum
Unit (ohms)
Examples
CAL:SHUN .01