EasyManuals Logo

Keysight Technologies 6811C Programming Guide

Keysight Technologies 6811C
315 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #67 background imageLoading...
Page #67 background image
Language Dictionary 3
Keysight Models 6811C, 6812C, and 6813C Programming Guide 67
CALibrate:SHUNt
CALibration:SHUNt
Keysight 6811C/6812C/6813C Option 020 only.
Allows you to enter the external current shunt value (in ohms). The external
current shunt is used for external current measurement. The programming range
is 1E-6 to 1E6. This parameter is non-volatile, so there is no *RST value. As with
other CAL commands, calibration must first be enabled with CAL:STATe, and the
value must be made permanent with CAL:SAVE.
Command syntax CALibrate:SHUNt <NRf+>
CALibration:SHUNt <NRf+>
Parameters 1E-6 to 1E6 | MINimum | MAXimum
Unit Ω (ohms)
Examples
CAL:SHUN .01

Table of Contents

Other manuals for Keysight Technologies 6811C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 6811C and is the answer not in the manual?

Keysight Technologies 6811C Specifications

General IconGeneral
BrandKeysight Technologies
Model6811C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals