3 Language Dictionary
122 Keysight Models 6811C, 6812C, and 6813C Programming Guide
CURRent:PROTection:STATe
This command enables or disables the overcurrent (OC) protection function. If the
overcurrent protection function is enabled and the ac source exceeds the
programmed level, then the output is disabled and the Questionable Condition
status register OC bit is set (refer to “Programming the Status Registers” on
page 259 in Chapter 5, "Programming Examples"). An overcurrent condition can
be cleared with OUTPut:PROTection:CLEar after the cause of the condition is
removed.
NOTE
Use OUTP:PROT:DEL to prevent momentary current limit conditions caused by
programmed output changes from tripping the overcurrent protection.
Command syntax [SOURce:]CURRent:PROTection:STATe <bool>
Parameters 0 | 1 | OFF | ON
*RST value OFF
Examples
CURR:PROT:STAT 1 CURR:PROT:STAT OFF
Query syntax [SOURce:]CURRent:PROTection:STATe?
Returned parameters 0 | 1
Related commands OUTP:PROT:CLE OUTP:PROT:DEL

Table of Contents

Other manuals for Keysight Technologies 6811C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 6811C and is the answer not in the manual?

Keysight Technologies 6811C Specifications

General IconGeneral
BrandKeysight Technologies
Model6811C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals