LIST:DWELl
This command sets the sequence of list dwell times. Each value represents the
time in seconds that the output will remain at the particular list step point before
completing the step. At the end of the dwell time, the output depends upon the
following conditions:
– If LIST:STEP AUTO has been programmed, the output automatically changes
to the next point in the list.
– If LIST:STEP ONCE has been programmed, the output remains at the present
level until a trigger sequences the next point in the list.
The order in which the points are entered determines the sequence in which they
are output when a list is triggered. Changing list data while a subsystem is in list
mode generates an implied ABORt.
LIST:DWELl:POINts?
This query returns the number of points specified in LIST:DWELl. Note that it
returns only the total number of points, not the point values.
Command syntax [SOURce:]LIST:DWELl <NRf+>{, <NRf+>}
Parameters 0 to 4.30349E5 | MINimum | MAXimum
Unit s (seconds)
Examples
LIST:DWEL 2.5, 1.5, .5
Query syntax [SOURce:]LIST:DWELl?
Returned parameters <NR3>{, <NR3>}
Related commands LIST:CURR LIST:FREQ LIST:TTLT LIST:VOLT
Query syntax [SOURce:]LIST:DWELl:POINts?
Returned parameters <NR1>
Examples
LIST:DWEL:POIN?