3 Language Dictionary
58 Keysight Models 6811C, 6812C, and 6813C Programming Guide
CALibrate:COUNt?
CALibration:COUNt?
Returns the calibration count of the ac source. Read and record the initial count
when you first receive your instrument. The value increments each time you store
the calibration constants, as such a complete calibration adds many counts. The
calibration count also increments whenever you save a calibration string, change
the calibration password, or override the calibration security. You can perform this
query regardless of whether the instrument is secured. This setting is non-volatile;
it will not be changed by power cycling, a factory reset (*RST), or an instrument
preset (SYSTem:PRESet).
:VOLTage
:AC Begin AC voltage calibration sequence
:AUXiliary
[:AC] Initiates calibration of the auxiliary output voltage
:DC Begin DC voltage calibration sequence
:EXTernal Initiates calibration of the external voltage measurement
:OFFSet Begin offset voltage calibration sequence
:PROTection Begin voltage protection calibration sequence
CALibrate | CALibration
Query syntax CALibrate:COUNt?
CALibration:COUNt?
Parameters None
Examples
CAL:COUN?
Returned parameters <NR1>
Related commands SYST:SEC:COUN