Keysight Models 6811C, 6812C, and 6813C Programming Guide 81
MEASure:ARRay:CURRent:HARMonic:PHASe?
FETCh:ARRay:CURRent:HARMonic:PHASe?
These queries return an array of harmonic phases of output current in degrees,
referenced to the positive zero crossing of the fundamental component.
The first value returned is the DC component (always returned as 0 degrees
phase), the second value is the fundamental frequency, and so on up to the 50th
harmonic. Harmonic orders can be measured up to the fundamental
measurement bandwidth of the measurement system, which is 12.6 kHz. Thus the
maximum harmonic that can be measured is dependent on the output frequency.
Any harmonics that represent frequencies greater than 12.6 kHz are returned as 0.
Query syntax MEASure:ARRay:CURRent:HARMonic:PHASe?
FETCh:ARRay:CURRent:HARMonic:PHASe?
Parameters None
Examples
MEAS:ARR:CURR:HARM:PHAS? FETC:ARR:CURR:HARM:PHAS?
Returned parameters 51 NR3 values
Related commands MEAS:ARR:VOLT:HARM:PHAS? MEAS:ARR:CURR:HARM?

Table of Contents

Other manuals for Keysight Technologies 6811C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 6811C and is the answer not in the manual?

Keysight Technologies 6811C Specifications

General IconGeneral
BrandKeysight Technologies
Model6811C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals