Keysight Models 6811C, 6812C, and 6813C Programming Guide 61
CALibrate:DATE
CALibration:DATE
Stores a message in calibration memory with a string input of 11 characters.
Storing a calibration message overwrites the previous message. You can perform
this query regardless of whether the instrument is secured. This setting is
non-volatile; it will not be changed by power cycling or *RST or SYSTem:PRESet.
CALibrate:IMPedance
CALibration:IMPedance
This command can only be used in calibration mode. It calibrates the output
impedance circuits. The command automatically performs the calibration.
CALibrate:SAVE is required to store the impedance constant in non-volatile
memory. CALibrate:IMPedance is a sequential command that takes several
seconds to complete.
Command syntax CALibrate:DATE <quoted_string>
CALibration:DATE <quoted_string>
Parameters <quoted_string>
Examples
CAL:DATE “14-Jun-2016”
Query syntax CALibrate:DATE?
CALibration:DATE?
Returned parameters <SRD>
Related commands CAL:STR
Command syntax CALibrate:IMPedance
CALibration:IMPedance
Parameters None
Examples
CAL:IMP
Related commands CAL:STAT CAL:SAVE