Keysight Models 6811C, 6812C, and 6813C Programming Guide 191
SYSTem:SECurity:COUNt?
Returns the security count of the ac source. The security count is incremented by
one every time you perform an action, other than calibration, that requires the ac
source to be unsecured. Be sure to read the security count upon receiving your
instrument.
SYSTem:SECurity
Sanitizes all user-accessible instrument memory except for the calibration
constants and reboots the instrument. This feature clears all user-defined state
information, measurement data, and user defined I/O settings such as the IP
address. This feature is not recommended for use in routine applications because
of the possibility of unintended data loss. The command is typically used before
removing an instrument from a secure area. The security count is incremented by
one each time you perform this command.
Query syntax SYSTem:SECurity:COUNt?
Parameters None
Examples
SYST:SEC:COUN?
Returned parameters <NR1>
Related commands CAL:COUN
Command syntax SYSTem:SECurity[:IMMediate]
Parameters None
Examples
SYST:SEC
Related commands SEC

Table of Contents

Other manuals for Keysight Technologies 6811C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 6811C and is the answer not in the manual?

Keysight Technologies 6811C Specifications

General IconGeneral
BrandKeysight Technologies
Model6811C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals