CURRent
This command sets the rms current limit of the specified output phase. If the
output current exceeds this limit, the output voltage amplitude is reduced until
the rms current is with the limit. The CL bit of the Questionable Status register
indicates that the current limit control loop is active. If the current protection state
is programmed on, the output latches into a disabled state when current limiting
occurs.
Command syntax [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <NRf+>
Parameters MINimum | MAXimum |
0 to 3.25 (6811C including all options)
0 to 6.50 (6812C including all options)
0 to 13.0 (6813C all options except Option 019)
0 to 16.0 (6813C Option 019)
Unit A (rms amperes)
*RST value MAXimum
Examples
CURR 5 CURR:LEV .5
Query syntax [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]?
Returned parameters <NR3>
Related commands CURR:PROT:STAT VOLT:RANG

Table of Contents

Other manuals for Keysight Technologies 6811C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 6811C and is the answer not in the manual?

Keysight Technologies 6811C Specifications

General IconGeneral
BrandKeysight Technologies
Model6811C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals