Language Dictionary 3
Keysight Models 6811C, 6812C, and 6813C Programming Guide 167
VOLTage:SLEW:TRIGgered
This command selects the slew rate that will be set during a triggered step or
pulse transient. A parameter of MAXimum or INFinity sets the slew to its
maximum possible rate. The SCPI representation for infinity is 9.9E37.
Command syntax [SOURce:]VOLTage:SLEW:TRIGgered <NRf+> | INFinity
Parameters 0 to 9.9E37 | MAXimum | MINimum | INFinity
Unit V (volts per second)
*RST value INFinity
Examples
VOLT:SLEW:TRIG 50 VOLT:SLEW:TRIG MAX
Query syntax [SOURce:]VOLTage:SLEW:TRIGgered?
Returned parameters <NR3>
Related commands VOLT:SLEW VOLT:SLEW:MODE

Table of Contents