CALibrate:PWM:RAMP
CALibration:PWM:RAMP
This command modulates the slope of voltage ramp driving the power output
stages. Varying the ramp affects the harmonic distortion of the output. The
argument is a number from 0 to 255. This command is only used during
manufacture or repair of the ac source.
CALibrate:SAVE
CALibration:SAVE
This command can only be used in calibration mode. It saves any new calibration
constants (after a current or voltage calibration procedure has been completed) in
non-volatile memory.
Command syntax CALCulate:PWM:RAMP <NRf>
CALibration:PWM:RAMP <NRf>
Parameters 0 through 255
Examples
CAL:PWM:RAMP 100
Query syntax CALibrate:PWM:RAMP?
CALibration:PWM:RAMP?
Returned parameters <NR1>
Related commands CAL:PWM:FREQ
Command syntax CALibrate:SAVE
CALibration:SAVE
Parameters None
Examples
CAL:SAVE
Related commands CAL:CURR CAL:VOLT CAL:STAT

Table of Contents

Other manuals for Keysight Technologies 6811C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 6811C and is the answer not in the manual?

Keysight Technologies 6811C Specifications

General IconGeneral
BrandKeysight Technologies
Model6811C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals