Language Dictionary 3
Keysight Models 6811C, 6812C, and 6813C Programming Guide 89
MEASure:CURRent:HARMonic:PHASe?
FETCh:CURRent:HARMonic:PHASe?
These queries return the phase angle of the Nth harmonic of output current,
referenced to the positive zero crossing of the fundamental component.
The parameter is the desired harmonic number. Queries sent with a value of 0
return the DC component. A value of 1 returns the fundamental output frequency.
Harmonic orders can be queried up to the fundamental measurement bandwidth
of the measurement system, which is 12.6 kHz. Thus the maximum harmonic that
can be measured is dependent on the output frequency. Any harmonics that
represent frequencies greater than 12.6 kHz are returned as 0.
MEASure:CURRent:HARMonic:THD?
FETCh:CURRent:HARMonic:THD?
These queries return the percentage of total harmonic distortion and noise in the
output current.
Query syntax MEASure:[SCALar]:CURRent:HARMonic:PHASe? <NRf>
FETCh:[SCALar]:CURRent:HARMonic:PHASe? <NRf>
Parameters 0 to 50
Examples
MEAS:CURR:HARM:PHAS? 3 FETC:CURR:HARM:PHAS? 1
Returned parameters <NR3>
Related commands MEAS:CURR:HARM? MEAS:CURR:HARM:THD?
Query syntax MEASure:[SCALar]:CURRent:HARMonic:THD?
FETCh:[SCALar]:CURRent:HARMonic:THD?
Parameters None
Examples
MEAS:CURR:HARM:THD? FETC:CURR:HARM:THD?
Returned parameters <NR3>
Related commands MEAS:CURR:HARM? MEAS:CURR:HARM:PHAS?