Language Dictionary 3
Keysight Models 6811C, 6812C, and 6813C Programming Guide 111
Sense Subsystem Commands
This subsystem controls the measurement current range, the data acquire
sequence, and the harmonic measurement window of the ac source.
For Keysight 6811C/6812C/6813C Option 020 only, the SENSe:NSELect and
SENSe:FREQuency:SOURce let you select a measurement source and the
frequency measurement channel for the Power Analyzer input. The
SENSe:SWEep:TINTerval, SENSe:WINDow, and other measurement time-based
commands apply to both measurement sources, regardless of which source is
selected.
Subsystem syntax
SENSe
:CURRent
:ACDC
:RANGe
[:UPPer] <n> Sets measurement current range
:FREQuency
:SOURce <source> Selects the frequency measurement source
:NSELect <n> Selects the measurement source that returns the data of a query
:SWEep
:OFFSet
:POINts <n> Define trigger points relative to the start of the digitizer data record
:TINTerval <n> Sets the digitizer sample spacing
:WINDow
[:TYPE] <type> Sets measurement window type (KBES | RECT)