Keysight Models 6811C, 6812C, and 6813C Programming Guide 197
Trigger Subsystem Commands
This subsystem controls the triggering of the ac source. Refer to “Triggering
Output Changes” on page 244 in Chapter 5, "Programming Examples" for an
explanation of the Trigger Subsystem. The INITiate commands control the
initialization of both the transient and measurement trigger systems.
For Keysight 6811C/6812C/6813C Option 020 only, TRIGger:ACQuire and other
measurement trigger commands apply to both measurement sources, regardless
of which source is selected.
Subsystem syntax
The trigger subsystem must first be enabled using the INITiate commands or no
triggering action will occur.
ABORt Resets the trigger system to the Idle state
INITiate
[:IMMediate] Initiates the system for one trigger
:SEQuence[1|3] Initiates a specific numbered sequence
:NAME <name> Initiates a specific named sequence (TRANsient | ACQuire)
:CONTinuous
:SEQuence[1] <bool> Sets continuous initialization
:NAME TRANsient <bool> Sets continuous initialization
TRIGger
[:SEQuence1 | :TRANsient]
[:IMMediate] Triggers the output immediately
:DELay <n> Sets the trigger delay time
:SOURce <source> Sets the trigger source (BUS | EXT | IMM)
:SEQuence2 | :SYNCronize
:SOURce <source> Sets the synchronous source (PHAse | IMMediate)
:PHASe <n> Sets the synchronous phase reference
:SEQuence3 | :ACQuire
[:IMMediate] Triggers the measurement immediately
:SOURce <source> Sets the trigger source (BUS | EXT | TTLT)
:SEQuence1
:DEFine TRANsient Sets or queries the SEQ1 name