3 Language Dictionary
60 Keysight Models 6811C, 6812C, and 6813C Programming Guide
CALibrate:DATA
CALibration:DATA
This command is only used in calibration mode. It enters a calibration value that
you obtain by reading an external meter. You must first select a calibration level
(with CALibrate:LEVel) for the value being entered. These constants are not stored
in non-volatile memory until they are saved with CALibrate:SAVE. If
CALibrate:STATE OFF is programmed without a CALibrate:SAVE, the previous
calibration constants are restored.
CALibrate:DATA?
CALibration:DATA?
Keysight 6812/6813C Option 026 only.
Returns the present value of the phase calibration. This command can only be
used during the auxiliary phase calibration procedure. It will return an error if it is
used at any other time.
Command syntax CALibrate:DATA <NRf>
CALibration:DATA <NRf>
Parameters <external_reading>
Examples
CAL:DATA 3222.3 MA CAL:DATA 5.000
Related commands CAL:STAT CAL:SAVE
Query syntax CALibrate:DATA?
CALibration:DATA?
Returned parameters <NR3>
Examples
CAL:DATA?

Table of Contents

Other manuals for Keysight Technologies 6811C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 6811C and is the answer not in the manual?

Keysight Technologies 6811C Specifications

General IconGeneral
BrandKeysight Technologies
Model6811C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals