3 Language Dictionary
200 Keysight Models 6811C, 6812C, and 6813C Programming Guide
INITiate:CONTinuous:SEQuence
INITiate:CONTinuous:NAME
These commands control the transient generator trigger system as follows:
– 1 or ON: Continuously initiates the transient trigger system.
– 0 or OFF: Turns off continuous triggering. In this state, the trigger system must
be initiated for each triggered event using INITiate:SEQuence.
INITiate:CONTinuous:SEQuence references the transient trigger sequence by its
number, while INITiate:CONTinuous:NAME references it by its name.
TRIGger
When the trigger subsystem has been initiated, the TRIGger command generates
a trigger signal regardless of the selected trigger source.
Command syntax INITiate:CONTinuous:SEQuence[1] <bool>
INITiate:CONTinuous:NAME TRANsient, <bool>
Parameters 0 | 1 | OFF | ON
Examples
INIT:CONT:SEQ ON INIT:CONT:NAME TRAN, 1
Related commands ABOR INIT:CONT TRIG *TRG
Command syntax TRIGger[:SEQuence1][:IMMediate]
TRIGger[:TRANsient][:IMMediate]
Parameters None
Examples
TRIG TRIG:TRAN TRIG:SEQ1:IMM
Related commands ABOR TRIG:SOUR TRIG:DEL TRIG:SYNC TRIG:SYNC:PHAS
INIT INIT:CONT *TRG WAI