EasyManuals Logo

Keysight Technologies 6811C Programming Guide

Keysight Technologies 6811C
315 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #132 background imageLoading...
Page #132 background image
3 Language Dictionary
132 Keysight Models 6811C, 6812C, and 6813C Programming Guide
FUNCtion:TRIGgered
This command selects the shape of the output voltage waveform when a triggered
step or pulse transient occurs. The parameters are as follows:
– SINusoid: A sine wave is the output.
– SQUare: A square wave is the output.
– CSINusoid: The output is a clipped sine wave. Both positive and negative peak
amplitudes are clipped at a value determined by the FUNCtion:CSINusoid
command.
– <table>: The output shape is described by one of the user-defined waveform
tables.
The maximum peak voltage that the ac source can output is 425 V
peak
. This
includes any combination of voltage, voltage offset, and function shape values.
Therefore, the maximum value that can be programmed depends on the
peak-to-rms ratio of the selected waveform. For a sine wave, the maximum
voltage that can be programmed is 300 V
rms
.
Command syntax [SOURce:]FUNCtion[:SHAPe]:TRIGgered <shape>
Parameters SINusoid | SQUare | CSINusoid | <table>
*RST value SINusoid
Examples
FUNC:TRIG SIN FUNC:TRIG TABLE1
Query syntax [SOURce:]FUNCtion[:SHAPe]:TRIGgered?
Returned parameters <CRD>
Related commands FUNC FUNC:MODE VOLT

Table of Contents

Other manuals for Keysight Technologies 6811C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 6811C and is the answer not in the manual?

Keysight Technologies 6811C Specifications

General IconGeneral
BrandKeysight Technologies
Model6811C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals