3 Language Dictionary
82 Keysight Models 6811C, 6812C, and 6813C Programming Guide
MEASure:ARRay:VOLTage?
FETCh:ARRay:VOLTage?
These queries return an array containing the instantaneous output voltage in
volts.
The output voltage and current are digitized whenever a measure command is
given or whenever an acquire trigger occurs. If digitization is caused by a measure
command, the time interval between samples is determined by the output
frequency. For frequencies greater than 45 Hz, the time interval is
25 microseconds. If digitization is caused by an acquire trigger, the time interval is
set by SENSe:SWEep:TINTerval, and the position of the trigger relative to the
beginning of the data buffer is determined by SENSe:SWEep:OFFSet:POINts.
Query syntax MEASure:ARRay:VOLTage[:DC]?
FETCh:ARRay:VOLTage[:DC]?
Parameters None
Examples
MEAS:ARR:VOLT? FETC:ARR:VOLT?
Returned parameters 4096 NR3 values
Related commands MEAS:ARR:CURR?