VOLTage:PROTection
This command sets the overvoltage protection (OVP) level of the ac source. If the
peak output voltage exceeds the OVP level, then the output is disabled and the
Questionable Condition status register OV bit is set (refer to “Programming the
Status Registers” on page 259 in Chapter 5, "Programming Examples"). An
overvoltage condition can be cleared with the OUTPut:PROTection:CLEar
command after the condition that caused the OVP trip is removed. The OVP
always trips with zero delay and is unaffected by the OUTPut:PROTection:DELay
command.
VOLTage:PROTection:STATe
This command enables or disables the overvoltage protection feature.
Command syntax [SOURce:]VOLTage:PROTection[:LEVel] <NRf+>
Parameters 0 to 500 | MAXimum | MINimum
Unit V (peak voltage)
*RST value MAXimum
Examples
VOLT:PROT 400 VOLT:PROT:LEV MAX
Query syntax [SOURce:]VOLTage:PROTection[:LEVel]?
Returned parameters <NR3>
Related commands OUTP:PROT:CLE OUTP:PROT:DEL
Command syntax [SOURce:]VOLTage:PROTection:STATe <bool>
Parameters 0 | 1 | OFF | ON
*RST value OFF
Examples
VOLT:PROT:STAT 1 VOLT:PROT:STAT ON
Query syntax [SOURce:]VOLTage:PROTection:STATe?
Returned parameters <NR3>
Related commands VOLT:PROT