3 Language Dictionary
92 Keysight Models 6811C, 6812C, and 6813C Programming Guide
Measurement Subsystem (Power) Commands
This subsystem programs the power measurement capability of the ac source.
Two measurement commands are available: MEASure and FETCh. MEASure
triggers the acquisition of new measurement data before returning a reading.
FETCh returns a reading computed from previously acquired data.
For Keysight 6811C/6812C/6813C Option 020 only, all standard ac source
MEASure or FETCh functions are available for the Main Output of the ac source as
well as the Power Analyzer input. The SENSe:NSELect command selects the
measurement source that will return data.
Subsystem syntax
MEASure:POWer?
FETCh:POWer?
These queries return the DC component of the power being sourced at the output
terminals in watts.
MEASure | FETCh
[:SCALar]
:POWer
[:DC]? Returns the DC component of power
:AC
[:REAL]? Returns real power
:APParent? Returns VA
:REACtive? Returns VAR
:PFACtor? Returns power factor
Query syntax MEASure:[SCALar]:POWer[:DC]?
FETCh:[SCALar]:POWer[:DC]?
Parameters None
Examples
MEAS:POW? FETC:POW?
Returned parameters <NR3>
Related commands MEAS:POW:AC?

Table of Contents

Other manuals for Keysight Technologies 6811C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 6811C and is the answer not in the manual?

Keysight Technologies 6811C Specifications

General IconGeneral
BrandKeysight Technologies
Model6811C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals