Keysight Models 6811C, 6812C, and 6813C Programming Guide 157
VOLTage:AUXiliary
Keysight 6812C/6813C Option 026 only
This command programs the AC rms output voltage level of the auxiliary output.
The auxiliary output waveform and frequency are always the same as the main
output. The auxiliary output is enabled along with the main output.
The maximum peak voltage of the auxiliary output is 36.8 V
peak
. This includes any
combination of voltage and function shape values. Therefore, the maximum value
that can be programmed depends on the peak-to-rms ratio of the selected
waveform. For a sine wave, the maximum voltage that can be programmed is
26.02 V
rms
.
Command syntax [SOURce:]VOLTage:AUXiliary[:LEVel][:IMMediate][:AMPLitude] <NRf+>
Parameters For sine waves: 0 to 26.02 | MAXimum | MINimum
Unit V (rms voltage)
*RST value 0
Examples
VOLT:AUX 12 VOLT:AUX:LEV 15
Query syntax [SOURce:]VOLTage:AUXiliary[:LEVel][:IMMediate][:AMPLitude]?
Returned parameters <NR3>
Related commands PHAS:AUX

Table of Contents

Other manuals for Keysight Technologies 6811C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 6811C and is the answer not in the manual?

Keysight Technologies 6811C Specifications

General IconGeneral
BrandKeysight Technologies
Model6811C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals