3 Language Dictionary
208 Keysight Models 6811C, 6812C, and 6813C Programming Guide
Bit configuration of Standard Event Status Enable register
*ESR?
This query reads the Standard Event Status Event register. Reading the register
clears it. The bit configuration of this register is the same as the Standard Event
Status Enable register (refer to “*ESE” on page 207). Refer to “Programming the
Status Registers” on page 259 in Chapter 5, "Programming Examples" for a
detailed explanation of this register.
Bit position 76543210
Bit name PON n.u. CME EXE DDE QYE n.u. OPC
Bit weight 128 32 16 8 4 1
n.u. = Not used
PON = Power-on
CME = Command error
EXE = Execution error
DDE = Device-dependent error
QYE = Query error
OPC = Operation complete
Command syntax *ESE <NRf>
Parameters 0 to 255
Power-on value Refer to *PSC
Examples *ESE 129
Query syntax *ESE?
Returned parameters <NR1>
Related commands *ESR? *PSC *STB?
Query syntax *ESR?
Parameters None
Returned parameters <NR1> (register value)
Related commands *CLS *ESE *ESE? *OPC

Table of Contents

Other manuals for Keysight Technologies 6811C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 6811C and is the answer not in the manual?

Keysight Technologies 6811C Specifications

General IconGeneral
BrandKeysight Technologies
Model6811C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals