166 Keysight Models 6811C, 6812C, and 6813C Programming Guide
VOLTage:SLEW
This command sets the slew rate for all programmed changes in the AC rms
output voltage level of the ac source. A parameter of MAXimum or INFinity sets
the slew to its maximum possible rate. The SCPI representation for INFinity is
9.9E37.
VOLTage:SLEW:MODE
This command determines how the output voltage slew rate is controlled during a
triggered output transient. The choices are as follows:
FIXed: The slew rate is unaffected by a triggered output transient.
STEP: The slew rate is programmed to the value set by
VOLTage:SLEW:TRIGgered when a triggered transient occurs.
PULSe: The slew rate is changed to the value set by VOLTage:SLEW:TRIGgered
for a duration determined by the pulse commands.
LIST: The slew rate is controlled by the voltage list when a triggered transient
occurs.
Command syntax [SOURce:]VOLTage:SLEW[:IMMediate] <NRf+> | INFinity
Parameters 0 to 9.9E37 | MAXimum | MINimum | INFinity
Unit V (volts per second)
*RST value INFinity
Examples
VOLT:SLEW 50 VOLT:SLEW INF
Query syntax [SOURce:]VOLTage:SLEW[:IMMediate]?
Returned parameters <NR3>
Related commands VOLT:SLEW:MODE VOLT:SLEW:TRIG
Command syntax [SOURce:]VOLTage:SLEW:MODE <mode>
Parameters FIXed | STEP | PULSe | LIST
*RST value FIXed
Examples
VOLT:SLEW:MODE STEP
Query syntax [SOURce:]VOLTage:SLEW:MODE?
Returned parameters <CRD>
Related commands VOLT:SLEW VOLT:SLEW:TRIG