Language Dictionary 3
Keysight Models 6811C, 6812C, and 6813C Programming Guide 213
*RCL
For standard Keysight 681xC series, this command restores the ac source to a
state that was previously stored in memory with a *SAV command to the specified
location. All states are recalled with the following exceptions:
– CAL:STATe is set to OFF
– The trigger system is set to the Idle state by an implied ABORt command (this
cancels any uncompleted trigger actions)
For Keysight 6812C/6813C Option 026 only, this command includes the auxiliary
output voltage and phase.
Recalling a previously stored state may place hazardous voltages at the ac
source output.
Rappeler un état précédemment mémorisé peut placer des tensions
dangereuses au niveau de la sortie de la source ac.
The device state stored in location 0 is automatically recalled at power turn-on
when the OUTPut:PON:STATE is set to RCL0.
Command syntax *RCL <NRf>
Parameters 0 to 15
Examples
*RCL 3
Related commands *PSC *RST *SAV