Loading...
Language Dictionary 3
Keysight Models 6811C, 6812C, and 6813C Programming Guide 201
TRIGger:DELay
This command sets the time delay between the detection of a trigger signal and
the start of any corresponding trigger action. After the time delay has elapsed, the
trigger is implemented unless the trigger system is also waiting for a sync signal
that has been specified by TRIGger:SYNChronous:PHASe.
TRIGger:SOURce
This command selects the trigger source for the first sequence in generating a
step, pulse, or list output as follows:
BUS: GPIB device, *TRG, or <GET> (Group Execute Trigger)
EXTernal: AC source's backplane Trigger In BNC
IMMediate: Trigger is generated as soon as the trigger system is initiated.
Command syntax TRIGger[:SEQuence1]:DELay <NRf+>
TRIGger[:TRANsient]:DELay <NRf+>
Parameters 0 to 4.30349E5 | MINimum | MAXimum
Unit s (seconds)
*RST value 0
Examples
TRIG:DEL .25 TRIG:DEL MAX TRIG:TRAN:DEL 1
Query syntax TRIGger[:SEQuence1]:DELay?
TRIGger[:TRANsient]:DELay?
Returned parameters <NR3>
Related commands ABOR TRIG TRIG:SOUR TRIG:SYNC TRIG:SYNC:PHAS
INIT INIT:CONT *TRG WAI
Command syntax TRIGger[:SEQuence1]:SOURce <CRD>
TRIGger[:TRANsient]:SOURce <CRD>
Parameters BUS | EXTernal | IMMediate
*RST value BUS
Examples
TRIG:SOUR BUS TRIG:TRAN:SOUR EXT
Query syntax TRIGger[:SEQuence1]:SOURce?
TRIGger[:TRANsient]:SOURce?
Returned parameters <CRD>
Related commands ABOR TRIG TRIG:DEL TRIG:SYNC TRIG:SYNC:PHAS
INIT INIT:CONT *TRG WAI

Table of Contents