VOLTage:SENSe:SOURce
VOLTage:ALC:SOURce
These commands select the source from which the output voltage is sensed. The
commands are interchangeable; they both perform the same function. The
following voltage sense sources can be selected:
– INTernal: This senses the voltage at the output of the power amplifier on the
inboard side of the output disconnect relay.
– EXTernal: This senses the output voltage at the rear panel voltage sense
terminals, which allows remote voltage sensing at the load.
Command syntax [SOURce:]VOLTage:SENSe:SOURce <source>
[SOURce:]VOLTage:ALC:SOURce <source>
Parameters INTernal | EXTernal
*RST value INTernal
Examples
VOLT:SENS:SOUR INT VOLT:ALC:SOUR EXT
Query syntax [SOURce:]VOLTage:SENSe:SOURce?
[SOURce:]VOLTage:ALC:SOURce?
Returned parameters <CRD>
Related commands VOLT:SENS:DET