Keysight Models 6811C, 6812C, and 6813C Programming Guide 57
Calibration Subsystem Commands
The commands in this subsystem allow you to do the following:
Enable and disable the calibration mode
Change the calibration password
Calibrate the current and voltage output levels, and store new calibration
constants in non-volatile memory.
Subsystem syntax
CALibrate | CALibration
:COUNt? Returns the calibration count of the ac source
:CURRent
:AC Begin AC current programming calibration sequence
:EXTernal Initiates calibration of the external current measurement
:MEASure Begin current measurement calibration sequence
:DATA <n> Input a calibration measurement
:DATA? Returns the present value of the phase calibration
:DATE <quoted_string> Input a calibration date
:IMPedance Begin output impedance calibration sequence
:LEVel <level> Advance to next calibration step (P1 | P2 | P3 | P4)
:PASSword <n> Set calibration password
:PHASe
:AUXiliary Initiates calibration of the auxiliary output phase
:PWM
:FREQuency <n> Trim pulse width modulator frequency
:RAMP <n> Trim pulse width modulator ramp
:SAVE Save new cal constants in non-volatile memory
:SECure
:CODE <quoted_string> Set calibration password
:STATe <bool>[, <n>] Unsecures or secures the instrument for calibration
:SHUNt <n> Allows you to enter the external current shunt value
:STATE <bool>[, <n>] Enable or disable calibration mode
:STRing <quoted_string> Stores a message in calibration memory