Language Dictionary 3
Keysight Models 6811C, 6812C, and 6813C Programming Guide 113
SENSe:FREQuency:SOURce
Keysight 6811C/6812C/6813C Option 020 only.
This command selects the frequency measurement source for the Power Analyzer
input. It only applies when the Power Analyzer input is active.
Accurate frequency measurements are critical when making all Harmonic
measurement calculations. Other measurement calculations also use frequency
measurement values, but they only slightly affect the accuracy of the
measurement. The ac source normally measures the frequency at the V SENSE
input channel. With this command you can select the I SENSE input channel as
the frequency measurement source for the Power Analyzer input.
Command syntax SENSe:CURRent:ACDC:RANGe[:UPPer] <NRf+>
Parameters 0 through 57.1342 | MINimum | MAXimum (all except Keysight 6811C)
0 through 28.5671 | MINimum | MAXimum (Keysight 6811C only)
Unit A (rms amperes)
*RST value MAX (high range)
Examples
SENS:CURR:ACDC:RANGE MIN
Query syntax SENSe:CURRent:ACDC:RANGe?
Returned parameters <NR3>
Related commands SENS:SWE:TINT MEAS:ARR
For Main Output measurements, the ac source uses the programmed output
frequency value when making measurement calculations.
Command syntax SENSe:FREQuency:SOURce <source>
Parameters VOLTage | CURRent
*RST value VOLTage
Examples
SENS:FREQ:SOUR CURR
Query syntax SENSe:FREQuency:SOURce?
Returned parameters <CRD>