3 Language Dictionary
162 Keysight Models 6811C, 6812C, and 6813C Programming Guide
VOLTage:OFFSet:SLEW:MODE
This command determines how the DC offset slew rate is controlled during a
triggered output transient. The choices are as follows:
– FIXed: The offset slew rate is unaffected by a triggered output transient.
– STEP: The offset slew rate is programmed to the value set by
VOLTage:OFFSet:SLEW:TRIGgered when a triggered transient occurs.
– PULSe: The offset slew rate is changed to the value set by
VOLTage:OFFSet:SLEW:TRIGgered for a duration determined by the pulse
commands.
– LIST: The offset slew rate is controlled by the voltage list when a triggered
transient occurs.
VOLTage:OFFSet:SLEW:TRIGgered
This command selects the DC offset slew rate that will be set during a triggered
step or pulse transient. A parameter of MAXimum or INFinity sets the slew to its
maximum possible rate. The SCPI representation for infinity is 9.9E37.
Command syntax [SOURce:]VOLTage:OFFSet:SLEW:MODE <mode>
Parameters FIXed | STEP | PULSe | LIST
*RST value FIXed
Examples
VOLT:OFFS:SLEW:MODE STEP
Query syntax [SOURce:]VOLTage:OFFSet:SLEW:MODE?
Returned parameters <CRD>
Related commands VOLT:OFFS:SLEW VOLT:OFFS:SLEW:TRIG
Command syntax [SOURce:]VOLTage:OFFSet:SLEW:TRIGgered <NRf+> | INFinity
Parameters 0 to 9.9E37 | MAXimum | MINimum | INFinity
Unit V (volts per second)
*RST value INFinity
Examples
VOLT:OFFS:SLEW:TRIG 50 VOLT:OFFS:SLEW:TRIG MAX
Query syntax [SOURce:]VOLTage:OFFSet:SLEW:TRIGgered?
Returned parameters <NR3>
Related commands VOLT:OFFS:SLEW VOLT:OFFS:SLEW:MODE