Keysight Models 6811C, 6812C, and 6813C Programming Guide 159
VOLTage:OFFSet
This command programs the DC output voltage level of the ac source.
The maximum peak voltage that the ac source can output is 425 V
peak
. This
includes any combination of voltage, voltage offset, and function shape values.
Therefore, the maximum value that can be programmed depends on the
peak-to-rms ratio of the selected waveform. For a sine wave, the maximum
voltage that can be programmed is 300 V
rms
.
NOTE
The OUTPut:COUPling must be set to DC to get non-zero DC output.
Command syntax [SOURce:]VOLTage:OFFSet[:IMMediate] <NRf+>
Parameters –425 to +425 | MAXimum | MINimum
Unit V (DC voltage)
*RST value 0
Examples
VOLT:OFFS 100
Query syntax [SOURce:]VOLTage:OFFSet[:IMMediate]?
Returned parameters <NR3>
Related commands VOLT:OFFS:MODE OUTP:COUP FUNC:SHAP