Language Dictionary 3
Keysight Models 6811C, 6812C, and 6813C Programming Guide 171
STATus:OPERation:NTRansition
STATus:OPERation:PTRansition
These commands set or read the value of the Operation NTR (Negative-Transition)
and PTR (Positive-Transition) registers. These registers serve as polarity filters
between the Operation Enable and Operation Event registers to cause the
following actions:
– When a bit in the Operation NTR register is set to 1, then a 1-to-0 transition of
the corresponding bit in the Operation Condition register causes that bit in the
Operation Event register to be set.
– When a bit of the Operation PTR register is set to 1, then a 0-to-1 transition of
the corresponding bit in the Operation Condition register causes that bit in the
Operation Event register to be set.
– If the same bits in both NTR and PTR registers are set to 1, then any transition
of that bit at the Operation Condition register sets the corresponding bit in the
Operation Event register.
– If the same bits in both NTR and PTR registers are set to 0, then no transition
of that bit at the Operation Condition register can set the corresponding bit in
the Operation Event register.
Setting a bit in the PTR or NTR filter can of itself generate positive or negative
events in the corresponding Operation Event register.
Command syntax STATus:OPERation:NTRansition <NRf+>
STATus:OPERation:PTRansition <NRf+>
Parameters 0 to 32767 | MAXimum | MINimum
Default value 0
Examples
STAT:OPER:NTR 32 STAT:OPER:PTR 1
Query syntax STATus:OPERation:NTRansition?
STATus:OPERation:PTRansition?
Returned parameters <NR1> (register value)
Related commands STAT:OPER:ENAB